Systems

Related Links

Goldfish Bluefish Analyzer Brochure

Instrument Size11.0"H x 14.3"W x 16.5"D (27.9 x 36.3 x 41.9 cm)

Fluidics bottle tray 10”H x 13.5"W x 4.5"D (25.4 x 34.2 x 11.4
cm)
Weight< 30 lbs (13.6 kg)
Emission detection
(using standard filter set
provided)

488 Excitation (Bluefish and Goldfish):
FL1: 533±15nm (FITC/GFP)
FL2: 585±20nm (PE/PI)
FL3: >670nm (PerCP-Cy5.5, PE-Cy5, PE-Cy7)

640 Excitation (Goldfish):
As above plus:
FL4: 675+12.5 nm (APC)
Optical alignmentFixed alignment, no maintenance required
Scatter resolutionResolves human peripheral blood erythrocytes, lymphocytes, monocytes, and granulocytes
Fluorescence linearity2.00+0.05% for chick erythrocyte nuclei (CEN)
Fluorescence precision< 3.0% CV for CEN
Max. event rate10,000 events/second
Signal processing24-bit A/D conversion
SoftwareIncellDx CFlow or IncellDx CFlow Plus Software